VENOUS OCCLUSION

BALLOON CATHETERS FOR ARTERIAL OR VENOUS OCCLUSION

OCCLUSION CATHETER

Used mainly to :
- occlude a vein or an artery by endovascular way,
- avoid the usage of a clamp,
- occlude rapidly in case of aortic hemorrhage.

The star shaped balloon allows a maximum development.


 
TECHNICAL CHARACTERISTICS
Reference
Size in F
Length in cm
color
Balloon volume in mL
Balloon diameter inflated in mm
65807
7
80
15,00
30
65808
8
80
38,00
40

 

Cathéter deux lumières

Cathéter d'occlusion temporaire permettant l'injection d'un produit en amont.
Les doubles lumières sont distales par défaut. Lumières proximales sur demande.



TECHNICAL CHARACTERISTICS
Reference
Size in F
Length in cm
color
Balloon volume in mL
Balloon diameter inflated in mm
265807
7
80
15,00
30
265808
8
80
38,00
40

USE

For single use only
Sterilized with ethylene oxide gas (ETO)

INFLATING THE BALLOON

Before use, inflate then deflate the balloon.
Do not exceed the indicated maximum balloon capacity, as this may cause a rupture of the latex balloon.
Air should not be used to inflate balloons, the balloon rupture during use could cause an air embolus. The balloon should be inflated with a sterile, pyrogen-free water or saline solution.

STORAGE

The catheters are supplied in sterile package.
They should be stored in a dry, cool place.
The packaging is designed to avoid the crushing of the catheter


For further information :

Phone : +33 (0)3.44.26.63.46
Fax : +33 (0)3.44.26.93.37

E-mail : contact@prodimed.com


 